Teledyne T802 Uživatelský manuál Strana 79

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 314
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 78
Teledyne API T802 Paramagnetic O
2
Analyzer Operation Manual
77
(The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory).
5.4.4. SETUP RNGE DIL: USING THE OPTIONAL DILUTION RATIO
FEATURE
This feature is a optional software utility that allows the user to compensate for
any dilution of the sample gas that may occur before it enters the sample inlet.
Typically this occurs in continuous emission monitoring (CEM) applications
where the sampling method used to remove the gas from the stack, dilutes the
sample.
To set up and use the dilution ratio option:
1. In the DIAG menu, use the 929 password and navigate to Factory Options
and enable the Dilution Ratio feature. (Press ENTER to save setting, and
then return to SETUP menu).
2. Select the reporting range mode and set the reporting range upper limit (see
Section 5.4.3).
07275B DCN6418
Zobrazit stránku 78
1 2 ... 74 75 76 77 78 79 80 81 82 83 84 ... 313 314

Komentáře k této Příručce

Žádné komentáře