Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. G, DCN 5247
Table 9-9-11: V/F Board Switch Settings
DAC # Sw 1 Sw 2 Sw 3 Sw 4 Sw 5 Sw 6 Sw 7
0 ON* OFF* OFF*
1 ON* OFF* OFF*
2 ON* OFF* OFF*
3 ON* OFF* OFF*
*Required settings
10 V output or non-isolated current loop
Switch State Comment
3 ON Use for non-isolated current loop or 10 V output
4 OFF Use for non-isolated current loop or 10 V output
5 OFF Use for non-isolated current loop or 10 V output
6 OFF Use for non-isolated current loop or 10 V output
5 V output or isolated current loop
Switch State Comment
3 OFF Use for isolated current loop or 5 V output
4 ON Use for isolated current loop or 5 V output
5 OFF Use for isolated current loop or 5 V output
6 OFF Use for isolated current loop or 5 V output
1 V output
Switch State Comment
3 OFF Use for 1 V output
4 OFF Use for 1 V output
5 ON Use for 1 V output
6 OFF Use for 1 V output
100 mV output
Switch State Comment
3 OFF Use for 100 mV output
4 OFF Use for 100 mV output
5 OFF Use for 100 mV output
6 ON Use for 100 mV output
9-44