Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
7.5 Test Measurements
All the test measurements which can be displayed by pressing the TEST button are also available
to the host computer via the RS-232 interface. The host computer should issue a request for a test
measurement, and then the CPU will send the current value of the test measurement to the RS-
232 output. The format of the test measurement message is:
T DDD:HH:MM IIII TEST MEASUREMENT<CRLF>
For example, the format of the DC Power Supply output in mV would be:
T 194:11:29 0000 DCPS= 2500 MV<CRLF>
To request a test measurement, the host must issue a command of the form:
T MEASUREMENT<CRLF>
For a summary of all test functions issue the command "T LIST". Table 7-8 lists the commands
and the corresponding test measurements which will be returned.
7-9