Calibration Procedures Model 360E Instruction Manual
152 05232 Rev B3
The default iDAS configuration records all calibration values in channel CALDAT as well as all calibration check
(zero and span) values in its internal memory. Up to 200 data points are stored for up 4 years of data (on
weekly calibration checks) and a lifetime history of monthly calibrations. Review these data to see if the zero
and span responses change over time. These channels also store the STABIL value (standard deviation of CO
concentration) to evaluate if the analyzer response has properly leveled off during the calibration procedure.
Finally, the CALDAT channel also stores the converter efficiency for review and documentation.
If your instrument has an O
2
sensor option installed that should be calibrated as well.