Operating Instructions Teledyne API – T101 Operation Manual
98
4.6.5. ELECTRICAL TEST
The electrical test function creates a current, which substitutes the PMT signal, and feeds
it into the preamplifier board. This signal is generated by circuitry on the pre-amplifier
board itself and tests the filtering and amplification functions of that assembly along with
the A/D converter on the motherboard. It does not test the PMT itself. The electrical test
should produce a PMT signal of about 2000 ±1000 mV.
To activate the electrical test press:
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > CAL SETUP
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CL K MORE EXIT
DIAG SIGNAL I / O
NEXT ENTR EXIT
DIAG ELEC RANGE = 500.000 PPB O2=X.XXX
<TST TST> EXIT
Pr ess NEXT until
DIAG ELECTRICAL TEST
PREV NE XT ENTR EXIT
Pre ss TST until…
DIAG ELEC PMT = 1732 MV SO2=X.X XX
<TST TST> EXIT
While th e ele ctr ical test is
activated, PMT should equal:
2000 mV ± 1000 mV
07266B DCN6485