Operating Instructions Teledyne API – T101 Operation Manual
84
4.6.2. ANALOG OUTPUT STEP TEST
This test can be used to check the accuracy and proper operation of the analog outputs.
The test forces all four analog output channels to produce signals ranging from 0% to
100% of the full scale range in 20% increments. This test is useful to verify the operation
of the data logging/recording devices attached to the analyzer.
To begin the Anog Output Step Test press:
SAMPLE RANGE = 500.0 PPB H2S =X XX. X
< TST TST > CAL SETUP
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X.X PRIM ARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
DIAG SIGNAL I / O
NEXT ENTR EXIT
Exit-Exit
returns to the
DIAG menu
SETUP X.X SECONDARY SETUP MENU
C OMM VARS DIAG EXIT
DIAG
NALOG OUTPUT
PREV NEXT ENTR EX IT
DIAG AOUT ANALOG OUTPUT
0% EXIT
Perform s
analog output
step test.
0% - 10 0%
DIAG AOUT ANALOG OUTPUT
[0%] EXIT
Pressi ng the b utton un de r “0%” whil e performing the test will pa use
the test at that l evel . Brackets will appe ar aro un d the value; exa mple:
[20%] Pressing the sam e button ag ain wi ll resu me the test.