Teledyne API – T101 Operation Manual Getting Started
53
3.6. INITIAL CALIBRATION
3.6.1. BASIC CALIBRATION PROCEDURE
The following three-step procedure assumes that the instrument does not have any of the
available zero/span (Z/S) or IZS valve options installed. Section 6 contains instructions
for calibrating instrume
nts with valve options.
The initial calibration should be carried out with the analyzer’s reporting range for
SINGLE range mode with a range span of 500 PPB (factory default settings for most
units). This will enable you to compare your results to the factory calibration.
STEP ONE: Set/verify the analog output reporting range of the T101:
SETUP X. X RANGE CONTROL MENU
MODE
SET U NIT EXIT
Press this button to se t
the analyzer for
SNGL
DUAL or AUT O ranges
SETUP X. X RANGE: 500.0 CONC
0 0 5 0 0 .0 ENTR EXIT
To chang e the val ue of the r epo rting
ra ng e span, en ter the nu mber by
pressing the button under each digit
until the expected value ap pears.
Pr ess th is but t on to se lect th e
concentration units of measure:
PPB, PPM, UGM, MGM
EXIT ignores the new setting and
returns to the RANGE CONTROL
MENU.
ENTR accepts the new setting and
returns to the
RANGE CONTROL MENU.
SETUP X. X
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > CAL SETUP
SETUP X. X R ANGE: 500.0 Con
0 0 0 5 0 .0 ENTR EXIT
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT