Operating Instructions Teledyne API – T101 Operation Manual
138
combined number of records would be exceeded. (For information regarding the
maximum number of channels, parameters, and records and how to calculate the file size
for each data channel, refer to the DAS manual downloadable from the TAPI website at
http://www.teledyne-api.com/manuals/ under Special Manuals). To set the number of
records for one channel from the front panel, press SETUP>DAS>EDIT>ENTR and the
following button sequence:
Edit Data Channel Menu
SETUP X.X 0) CONC: ATIMER, 2, 900
PREV NEXT INS DEL EDIT PRNT EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X NAME:CONC
<SET SET> EDIT PRINT EXIT
SETUP X.X NUMBER OF RECORDS:000
<SET SET> EDIT PRINT EXIT
SETUP X.X REPORT PERIODD:DAYS:0
0 0 0 0 0 ENTR EXIT
ENTR accepts the new
setting and returns to the
previous m enu.
EXIT ignores the new setting
and returns to the previous
menu.
Toggle buttons to set
number of records
(1-99999)
SETUP X.X EDIT RECOPRDS (DELET DATA)
YES NO
NO returns to the
p
revious menu.YES will delete all data
in this channel.
Press SET> key until
From the DATA ACQUISITION menu
(see Section 6.1 2.2.2)
07266B DCN6485