Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485)
A-22
WPMTTEMP PMT TEMP WARNING PMT temperature outside of warning
limits specified by PMT_SET variable.
WDARKCAL DARK CAL WARNING Dark offset above limit specified by
DARK_LIMIT variable.
WHVPS HVPS WARNING High voltage power supply output
outside of warning limits specified by
HVPS_SET variable.
WDYNZERO CANNOT DYN ZERO Contact closure zero calibration
failed while DYN_ZERO was set to
ON.
WDYNSPAN CANNOT DYN SPAN Contact closure span calibration
failed while DYN_SPAN was set to
ON.
WREARBOARD REAR BOARD NOT DET Rear board was not detected during
power up.
WRELAYBOARD RELAY BOARD WARN Firmware is unable to communicate
with the relay board.
WFRONTPANEL FRONT PANEL WARN Firmware is unable to communicate
with the front panel.
WANALOGCAL ANALOG CAL WARNING The A/D or at least one D/A channel
has not been calibrated.
1
The name is used to request a message via the RS-232 interface, as in “T BOXTEMP”.
2
Engineering software.
3
Current instrument units.
4
T101, M101E.
5
Triple-gas option.
6
T108, M108E.
7
Concentration alarm option.
8
T108U, M108EU.
9
Optional.